Flags
6-23
Flag
Triggering Criteria
CBC
CBC /
DIFF
CBC / DIFF
/ RETIC
CBC /
RETIC
RETIC
PX-NO (NX) 1.
Perox % Noise > 60%
2.
Flag is not set if there is
WBC agreement (no
WBC-CE flag) and
[(%NEUT + %EOS) -
%PMN] is between 0
and 7.5.
PX-PL (PX)
Perox light intensity < 90
PX-SAT
(XS)
1.
Perox % Saturation > 10
2.
Flag is not set if there is
WBC agreement (no
WBC-CE flag).
PXTO (TX)
1.
Perox chamber
temperature is not
between 58°C and
72.1°C
2.
Flag is not set if there is
WBC agreement (no
WBC-CE flag) and
[(%NEUT + %EOS) -
%PMN] is between 0
and 7.5.
RBCIFR
(RR)
RBC Flatness > 3.2
RTCint (CT)
Platelet threshold below
channel 5
RTCADA
(CA)
1.
Absorption mode for
Gaussian fit is not
between channels 6 and
30
2.
SD of RETIC Abs
histogram over 80% of
height > 1.4
3.
Measured mean
absorption minus the
theoretical absorption
> 30%
RTCIFR
(CR)
Retic Flatness > 3.2
Summary of Contents for ADVIA 2120