7-68
Status Line messages
2 = arithmetic mean
Patient QC Formula Not Defined – Parameterization – Calcul
This error occurs when there is no value defined for the FSE parameter
Calculs_PatientQC_Formula. This parameter determines the formula used to
calculate the moving average. To define the parameter, contact your Siemens
Field Service Representative.
The possible settings for the parameter are:
0 = parameter not active
1 = Bull’s Algorithm
2 = arithmetic mean
Patient Result Database Is Full
The patient result database has reached maximum capacity, causing the
autosampler to stop and transition the system state to "Aspiration Paused."
Corrective Action
Increase available patient result database space by deleting data from the sample
results file. Use Customize, System Setup, Tools Modify, End of Day.
IMPORTANT
Deleting files permanently erases data. If you need an electronic copy of the data,
back up the files before deleting.
Patient Result Database Is Nearly Full
The patient result database has reached the level specified as the warning level,
causing the autosampler to stop and transition the system state to "Ready to
Run."
Corrective Action
Increase available patient result database space by deleting data from the sample
results file. Use Customize, System Setup, Tools Modify, End of Day.
IMPORTANT
Deleting files permanently erases data. If you need an electronic copy of the data,
back up the files before deleting.
Summary of Contents for ADVIA 2120
Page 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Page 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Page 46: ...3 12 Daily Routine ...
Page 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Page 269: ...7 100 Status Line messages ...
Page 341: ...8 72 Methods ...