Status Line messages
7-77
i
Select
Shut Down ADVIA
at the Log On/Off
window.
ii
Select
Off
at the analyzer touchpad.
iii
Select
CTRL+ALT+DELETE
and log back onto
the system to restart the software.
iv
Restart the analyzer by selecting
On
at the
analyzer touchpad.
2
Pressure out of
range.
Verify pressure readings and adjust if necessary.
3
Faulty power
supply, fuse,
cable,
Pressure/Switc
h Node Board,
or Can Bus
Scrambler
Board.
Call Siemens Service for assistance.
Printer Busy - Printing Aborted
The printer was busy when a printout was requested.
Corrective Action
Restart the printout when the printer is free.
Printer Not Defined
The printer is not defined in the Port Configuration window.
Corrective Action
Correct or enter the printer definition in the Port Configuration window. Select
Customize menu, System Setup, Tools Modify, Port Configuration.
Pr
i
nter Not Ready - Impression Aborted
A printer problem has occurred during background printing of a results report.
No message box is displayed. The printout will resume as soon as the printer is
available.
Corrective Action
Verify printer connections and operation.