Flags
6-39
Normal
Perox Cytogram
PX-SAT
Perox Cytogram
1 Saturation area
The PX-SAT (XS) flag is not triggered if WBCB and WBCP agree within
specified limits (no WBC-CE flag).
The WBCP count does not include events from the Saturation area.
Results Flagged
WBCP, %NEUT, #NEUT, %LYMPH, #LYMPH, %MONO, #MONO, %EOS,
#EOS, %LUC, and #LUC
Corrective Action
Multiple occurrences of this flag, especially for consecutive samples, can indicate
a system problem. However, isolated instances of this flag are usually sample
related.
The following list of sample-related causes may not contain all conditions that
could cause this flag. There is no intention to associate the flag with specific
diagnoses.
System Related
Sample Related
1.
Check perox reagents.
2.
Check peroxidase hydraulics and sample
delivery.
3.
Check delivery of SHEATH/RINSE.
4.
Check PEROX SHEATH delivery.
5.
Check the pressure and vacuum readings.
6.
Check the perox reaction chamber temperature.
7.
Check perox flowcell alignment.
8.
Check perox gains.
•
Immature
granulocytes
•
Severe toxic
granulation
•
Promyelocytic
leukemia