Welcome to the ADVIA 2120/2120i Hematology System
1-11
The sample pumps (
1
)
dispense an exact amount
of reacted sample (
2
) from
the appropriate reaction
chamber (
3
) into the
flowcells (
4
).
The sheath diaphragm
pumps (
5
) push the sheath
(
6
) through the sheath
filters (
7
) directly into the
flowcells.
The sheath pumps (
8
) pull
the PEROX SHEATH or
SHEATH / RINSE and the
reacted sample stream (
9
)
through the flowcells for
analysis.
The analyzed sample and
sheath are sent to waste
(
10
) and the lines,
flowcells, and reaction
chambers are washed and
rinsed.