ADVIA Autoslide Slide Maker Stainer
10-21
3.
Partial draining of May-Grünwald Stain (Stain 1) (AN2).
4.
Dispensing Buffer to dilute May-Grünwald Stain (Stain1) (DN3).
5.
Draining (AN3).
6.
Dispensing diluted Giemsa Stain (Stain 2) (DN3).
7.
Draining (AN7).
8.
Dispensing rinsing solution (distilled water) (DN7).
9.
Draining (AN7).
10.
Dispensing rinsing solution (distilled water) (DN7).
11.
Draining (AN7).
12.
Draining (AN4).
13.
Dispensing methanol (DN4).
14.
Slide removal
Needle Position
Action
DN1 (Undil. Stain 1)
32
1
DN2 (Buffer)
8
1
DN3 (Dil. Stain 2)
13
1
DN4 (Methanol)
31
0
DN7 (Rinse)
18
2
AN2 8
1
AN3 13
1
AN4 31
1
AN7 18
3
Methanol Volume (mL) 1.6
Stain Volume (mL)
3.6
Dilution Ratio (%)
5
Duration Long
Principle of the Procedure
The 2 staining mixtures used by this protocol have very distinct proprieties. The
May-Grünwald (STAIN 1) stains acidophilic sites of cells and the neutrophilic
granulations of leukocytes, whereas the Giemsa (STAIN 2) stains the cytoplasm
of the monocytes and lymphocytes, as well as the nuclear chromatin. The stained
smear is examined microscopically and cells are then manually differentiated and
quantified.
Cell Type
Cell Component
Color
Neutrophils Nuclei
Deep
purple
Summary of Contents for ADVIA 2120
Page 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Page 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Page 46: ...3 12 Daily Routine ...
Page 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Page 269: ...7 100 Status Line messages ...
Page 341: ...8 72 Methods ...