ADVIA Autoslide Slide Maker Stainer
10-45
9.
Close the tape holder plate.
10.
At the Autoslide Control
window, select
Start
to tighten the tape.
Replacing the Printer Ribbon
Materials required: New print cartridge
Time: 2 minutes
Analyzer mode: Standby
To replace the print ribbon
CAUTION
Make sure
all
slides are processed and transferred to the output queue,
before
turning off the Autoslide. Turning off the autoslide while slides are processing
can cause unreliable results and possibly damage Autoslide components.
1.
Turn off the Autoslide.
2.
Open the lower front cover and remove the slide rack loader to gain easy
access to the printer ribbon.