Regulatory Information
WBC DIFF Method Data
The sample stability, expected ranges, and performance characteristics for the WBC DIFF method are listed in the following Table.
Sample Stability
Expected
Imprecision
Correlation Data
Room Temp
Refrigerated
Values
Mean
SD
CV%
y=
r
Sy.x
%NEUT
36 h
72 h
40 to 77
55.0
0.9
1.6
1.02x - 0.6
0.997
1.6
%LYMPH
36 h
72 h
16 to 44
31.3
0.9
2.9
1.00x + 0.8
0.997
1.7
%MONO
72 h
72 h
4 to 9
7.2
0.5
6.9
0.85x - 0.3
0.943
0.8
%EOS
8 h
72 h
1 to 7
3.4
0.3
8.8
0.87x + 0.2
0.979
0.4
%BASO
72 h
56 h
0 to 1
0.5
0.1
20.0
0.67x + 0.0
0.772
0.2
%LUC
72 h
72 h
1 to 4
2.5
0.4
16.0
0.92x + 0.6
0.994
0.7