
9-52
Regulatory Information
5.
In certain abnormal specimens (specifically lymphoid disorders and
leukemias), erroneously elevated basophil counts may be observed.
Expected Values
The range of expected values can vary depending on age, sex, diet, and location.
It is recommended that each laboratory establish its own range of expected
values.
The following range of expected values is based upon the duplicate assay of 60
whole-blood specimens obtained from an adult population of assumed healthy
individuals.
Parameter
Low Value High Value
%NEUT 40 77
%LYMPH 16
44
%MONO 4 9
%EOS 1 7
%BASO 0 1
%LUC 1 4
NOTE:
Because of numerical rounding, the Diff results should total 100%
±
0.2%,
except when %BASO is >5% and the Baso Count Suspect (B-SUSP, BC) flag is
triggered. In cases where the total is greater than 100%
±
0.2%, the amount in
excess of 100% is due to %BASO.
Performance Characteristics: Imprecision
A total of 18 runs were performed using multiple samples.
The following estimates of imprecision were obtained by taking 25 aspirations
from each whole-blood sample. Each of the sample selectivity modes was tested
by sampling from opened tubes and with the manual and automated closed-tube
samplers.
Parameter Mean Standard
Deviation
CV (%)
%NEUT 55.0
0.9
1.6
%LYMPH 31.3
0.9
2.9
%MONO 7.2
0.5
6.9
%EOS 3.4
0.3
8.8
%BASO 0.5
0.1
20.0
%LUC 2.5
0.4 16.0
Summary of Contents for ADVIA 2120
Page 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Page 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Page 46: ...3 12 Daily Routine ...
Page 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Page 269: ...7 100 Status Line messages ...
Page 341: ...8 72 Methods ...