6-4
Flags
Flag Triggering
Criteria CBC CBC
/
DIFF
CBC / DIFF
/ RETIC
CBC /
RETIC
RETIC
LPLT
%LPLT > 10% PLT
LS
BASO d/D < 0.15 and
%NEUT
≥
30%
MACRO %MACRO
≥
2.5%
MICRO %MICRO
≥
2.5%
MPO-D
(MO)
[%PMN
−
(%NEUT +
%EOS)]
≥
25, no NRBC
flag and a valid MN-PMN
valley (d/D
≥ 0.15)
NRBC 1.
WBCu
≥
to 3000 cells
/ µL and % nRBC
≥
2.0%
(# nRBC / 100WBC)
or
2. # nRBC
≥
200 cells / µL
PLT-CLM
(NW)
Clumps Count > 150
RBCF
RBCF > 100,000 cells/µL
RBCG
RBCG > 100,000 cells/µL
Anisocytosis (ANISO)
Definition
The Anisocytosis flag is triggered if variation in RBC volume (RDW) is equal to
or greater than 16%. The Red Cell Distribution Width (RDW) parameter is the
coefficient of variation of the cell volume distribution on the RBC volume
histogram.
Default trigger values for the three severity levels are:
+
RDW = 16.0% to 17.9%
++
RDW = 18.0% to 22.0%
+++
RDW > 22.0%