Status Line messages
Corrective Action
The alarm counter is automatically reset after reagents are replenished and the
Reagent Installation window is updated. Go to Logs, Reagent Log, Reagent
Installation.
RBC Sample Pump Error - Reset Required
RBC sample syringe pump not in home position, failed to respond, or a
transmission error occurred.
Possible Cause
1
RBC sample
syringe pump
not moving or
not in home
position.
Corrective Action
a
Reset the analyzer by selecting
Utilities, Exerciser,
Indicators, Analyzer Reset
.
b
After about 1 minute, if the status line does not
display Ready to Run and the green ready-to-aspirate
light is still off, follow these steps to exit the ADVIA
2120/2120i software, turn off the analyzer and then
restart.
i
Select
Shut Down ADVIA
at the Log On/Off
window.
ii
Select
Off
at the analyzer touchpad.
iii
Select
CTRL+ALT+DELETE
and log back onto
the system to restart the software.
iv
Restart the analyzer by selecting
On
at the
analyzer touchpad.
2
RBC sample
syringe
incorrectly
assembled.
Follow the steps in Replacing the Syringe Plungers.
3
Faulty power
supply, fuse,
cable, pump
assembly, Can
Bus Scrambler
Board, or Dual
Servo Pump
Control Board.
Call Siemens Service for assistance.
RBC Sheath Pump Error - Reset Required
RBC Sheath syringe pump not in home position, failed to respond, or a
transmission error occurred.
Summary of Contents for ADVIA 2120
Page 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Page 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Page 46: ...3 12 Daily Routine ...
Page 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Page 269: ...7 100 Status Line messages ...
Page 341: ...8 72 Methods ...