9-30
Regulatory Information
Parameter
Low Value
High Value
MCV (fL)
76
100
MCH (pg)
24
31
MCHC (g/dL)
28
34
CH (pg)
24
35
CHCM (g/dL)
29
34
RDW (%)
12
15
HDW (g/dL)
1.9
3.0
PLT (10³/µL)
140
360
MPV (fL)
7
9
Performance Characteristics: Imprecision
A total of 18 runs were performed using multiple samples.
The following estimates of imprecision were obtained by taking 25 aspirations
from each whole-blood sample. Each of the sample selectivity modes was tested
by sampling from opened tubes, and with the manual and automated closed-tube
samplers.
Parameter Mean
Standard
Deviation
CV
(%)
WBC (10³/µL)
5.91
0.16
2.7
RBC (10
6
/µL) 4.82
0.05
1.0
HGB (g/dL)
13.3
0.11
0.8
MCV (fL)
81.0
0.3
0.4
CH (pg)
26.5
0.05
0.2
CHCM (g/dL)
30.6
0.14
0.5
RDW (%)
12.3
0.07
0.6
HDW (g/dL)
2.56
0.02
0.8
PLT (10³/µL)
256
6.9
2.7
MPV (fL)
8.1
0.11
1.4
Performance Characteristics: Correlation Data
The performance of the ADVIA 120 Hematology System was compared with the
performance of the Technicon H
●
3 RTC System (x) using whole blood from 60
normal donors and 70 hospital donors with each sample run in duplicate (n =
260).
Summary of Contents for ADVIA 2120
Page 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Page 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Page 46: ...3 12 Daily Routine ...
Page 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Page 269: ...7 100 Status Line messages ...
Page 341: ...8 72 Methods ...