7-64
Status Line messages
No Working Buffer Initialized
This error can occur during the system preparation (initialization) process.
Corrective Action
Restart the ADVIA 2120/2120i software. If the error recurs, call Siemens Service
for assistance.
Noise BASO - Alarm
The number of consecutive occurrences of this condition has met the criterion
specified for an alarm message. The specified alarm criterion is user definable in
the Customize Menu, System Setup, Alarm/Stop Criteria. The counter associated
with the Alarm Criteria is automatically reset.
Noise BASO - Stop
The number of consecutive occurrences of this condition has met the criterion
specified for a stop message. The autosampler halts at the same time the stop
message occurs. The specified stop criterion is user definable in the Customize
Menu, System Setup, Alarm/Stop Criteria. The counter associated with the Stop
Criteria is reset when the Start/Stop button is selected.
Noise PEROX - Alarm
The number of consecutive occurrences of this condition has met the criterion
specified for an alarm message. The specified alarm criterion is user definable in
the Customize Menu, System Setup, Alarm/Stop Criteria. The counter associated
with the Alarm Criteria is automatically reset.
Noise PEROX - Stop
The number of consecutive occurrences of this condition has met the criterion
specified for a stop message. The autosampler halts at the same time the stop
message occurs. The specified stop criterion is user definable in the Customize
Menu, System Setup, Alarm/Stop Criteria. The counter associated with the Stop
Criteria is reset when the Start/Stop button is selected.
Non WBC Interference - Alarm
The number of consecutive occurrences of this condition has met the criterion
specified for an alarm message. The specified alarm criterion is user definable in
the Customize Menu, System Setup, Alarm/Stop Criteria. The counter associated
with the Alarm Criteria is automatically reset.
Summary of Contents for ADVIA 2120
Page 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Page 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Page 46: ...3 12 Daily Routine ...
Page 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Page 269: ...7 100 Status Line messages ...
Page 341: ...8 72 Methods ...