5-38
Troubleshooting the Analyzer
b.
Visually check for bubbles in the open-tube sample line.
c.
Run enough controls to verify analyzer performance. If control results
are not acceptable, check that the probe has been properly installed. If no
problem is found, recalibrate the affected channel.
Replacing the Perox Lamp
WARNING
To prevent injury to eyes from shattered glass, wear safety glasses when handling
tungsten-halogen lamps.
To avoid severe burns, allow sufficient time for the lamp to cool. After the
analyzer is turned off, wait at least 10 minutes.
CAUTION
Fingerprints on the lamp could cause severe degradation in lamp performance.
DO NOT touch the lamp with your fingers.
Material required: Replacement lamp, PN 065-B075-01
Time: Replacement - 20 minutes
Adjustment and checkout - 20 minutes
Analyzer mode: Off, wait at least 10 minutes.
ELECTRICAL WARNING
To avoid exposure to shock hazards and/or damage to the instrument while
performing this procedure, power off the analyzer before proceeding.
To replace the perox lamp (
1
)
IMPORTANT
Make sure that the power is off before disconnecting the lamp.
1.
Disconnect the lamp power connector (
2
) in the back of the perox optics
assembly by squeezing the sides of the red connector and pulling it apart.
FRONT OF ANALYZER
Summary of Contents for ADVIA 2120
Page 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Page 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Page 46: ...3 12 Daily Routine ...
Page 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Page 269: ...7 100 Status Line messages ...
Page 341: ...8 72 Methods ...