6-22
Flags
Flag
Triggering Criteria
CBC
CBC /
DIFF
CBC / DIFF
/ RETIC
CBC /
RETIC
RETIC
HGBIFR
(HR)
HGB Sample Flatness
> 1000
HGB-PL
(PH)
HGB baseline is not
between 2.5 and 4.1
LAS-PL(PL)
Laser light intensity < 150
NRCELL
(NC)
WBCP / WBCB > 1.1
NR-LPD
(NL)
1. BASO % Noise > 2
2. BASO % Noise x WBCB
> 10
NRLPLT
(NP)
LPLT > 40 x 10
3
/
μ
L
NRPXNV
No valley between the
NRBC and lymphocyte
populations and a
reportable NRBC count.
PLT-NO
(NT)
1.
LPLT and ANISO flags
are triggered (any
severity level
2.
LPLT and MICRO flags
are triggered (any
severity level).
PLTORN
(OT)
Events in the Platelet
Origin Noise area are > 2%
of the total events the PLT
cytogram
PX-NV (VX) 1.
(PEROX d/D) < 0.15
2.
Flag is not set if there is
WBC agreement (no
WBC-CE flag) and
[(%NEUT + %EOS) -
%PMN] is between 0
and 7.5.
PXIFR (XR)
Perox Flatness > 3.2