Status Line messages
7-49
F
File Does Not Exist
A patient or control result printout has been selected in the Routine Parameters
window. But one of the files, Report.par, Prevalid.par, Print.par, or QCreport.par,
is not defined.
Corrective Action
Restore dictionaries in Other Utilities window.
File Not Found: C:\Wst\Oen.cnf
The default workorder screen configuration file is missing.
Corrective Action
Continue to create workorders in Order Entry as needed. To restore the Oen.cnf
file, select Other Utilities and restore dictionaries.
File Not In Modifiable File List - Updating
This message appears when a file is created through the Format window. The file
is then stored in the database.
Corrective Action
No action is required.
G
GETEQUAL
This control is not defined in the Control Dictionary.
Corrective Action
Define the control in the Control Dictionary.
GlobalLock or GlobalUnlock
The selected dispositions cannot be stored because the sample record is full.
Corrective Action
Reduce the number of dispositions selected or cancel the selections and create a
new workorder through the Order Entry window.
Summary of Contents for ADVIA 2120
Page 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Page 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Page 46: ...3 12 Daily Routine ...
Page 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Page 269: ...7 100 Status Line messages ...
Page 341: ...8 72 Methods ...