Troubleshooting the Analyzer
5-15
5.
Release the flowcell by loosening the release knob located on the right of the
flowcell adjustment assembly
Flowcell release knob (
1
)
Front of analyzer
6.
Hold the flowcell by the red threaded fitting located at the top of the flowcell,
then gently lift the flowcell out of the optics assembly.
CAUTION
To avoid getting fingerprints on the glass windows of the flowcell, always
hold the flowcell by the metal slides.
Cleaning the Flowcells off the Analyzer: Step 1
Removing the RBC/BASO/Retic Flowcell
Before you begin the cleaning of the RBC/baso/retic flowcell, read all laser
safety precautions.