Status Line messages
7-45
Compressor Timeout
During system startup or when exiting standby, the compressor has failed to
reach the acceptable pressure or vacuum levels within the allotted time.
Possible Cause
1
System
restarted too
quickly after
being turned
off
Corrective Action
Manually vent the waste
container by
disconnecting the waste
container vacuum line
(1), waiting for the
vacuum to dissipate,
and then reconnecting
the line.
1
2
Vacuum line
attached to the
waste container
is disconnected
or crimped
a
Reconnect or remove any crimps from the waste
container vacuum lines.
b
At the analyzer touchpad, select
Off
.
c
Wait at least 1 minute and select
On
to restart the
analyzer.
3
Vacuum pump
filter
(vacushield) is
clogged
Replace the vacuum pump filter.
4
Compressor
failure
Call Siemens service for assistance.
Control Out of Range - Alarm
One or more control samples have yielded out of range test results. The number
of consecutive occurrences of this condition has met the criterion specified for an
alarm message. The specified alarm criterion is user definable in the Customize
Menu, System Setup, Alarm/Stop Criteria.
Control Out of Range - Stop
One or more control samples have yielded out of range test results. The number
of consecutive occurrences of this condition has met the criterion specified for a
stop message. The autosampler halts when the stop message occurs. The
specified stop criterion is user definable in the Customize Menu, System Setup,
Alarm/Stop Criteria.
Could Not Attempt Location of a Workorder, Exception Occurred
A sample workorder could not be located because of a software error.
Corrective Action
Summary of Contents for ADVIA 2120
Page 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Page 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Page 46: ...3 12 Daily Routine ...
Page 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Page 269: ...7 100 Status Line messages ...
Page 341: ...8 72 Methods ...