ADVIA Autoslide Slide Maker Stainer
10-41
Materials required:
•
Methanol
•
Laboratory tray
•
Paper towels
Time: 15 minutes
Analyzer mode: Standby
NOTE
It is recommended that the stainer wells be replaced once a year. Check the
Replaceable Parts section of the Operator's Guide for ordering information.
TOXIC
Methanol is toxic. To avoid serious irreversible effects do not inhale, do not
contact with skin and do not swallow. Wear suitable protective clothing and
gloves. In case of accident or if you feel unwell, seek medical advice
immediately.
FLAMMABLE
Methanol is flammable. Keep away from sources of ignition - no smoking.
CAUTION
Make sure
all
slides are processed and transferred to the output queue,
before
turning off the Autoslide. Turning off the autoslide while slides are processing
can cause unreliable results and possibly damage Autoslide components.
1.
Put the analyzer in Standby mode and turn off the Autoslide.
2.
Slide open the stainer tray door.
3.
While manually rotating the stainer plate, remove the stainer wells by gently
sliding each out of the stainer.