7-14
Status Line messages
autoRETIC Reagent Empty - Stop
The number of consecutive occurrences of this condition has met the criterion
specified for a stop message. The reagent is empty, and the autosampler has
halted. The specified alarm criterion is user definable in the Customize Menu,
System Setup, Alarm/Stop Criteria, Reagent Conditions.
Corrective Action
The stop counter is automatically reset after reagents are replenished and the
Reagent Installation window is updated. Go to Logs, Reagent Log, Reagent
Installation. Select Start/Stop to continue the run.
autoRETIC Reagent Expired
Present date is beyond the autoRETIC reagent expiration date, as determined by
encoded reagent container barcode label or as calculated from installation date
and open container stability.
Corrective Action
Replenish expired reagents and update the Reagent Installation window. Go to
Logs, Reagent Log, Reagent Installation.
autoRETIC Reagent Low - Alarm
The number of consecutive occurrences of this condition has met the criterion
specified for an alarm message, indicating that the reagent level is low. The
specified alarm criterion is user definable in the Customize Menu, System Setup,
Alarm/Stop Criteria, Reagent Conditions.
Corrective Action
The alarm counter is automatically reset after reagents are replenished and the
Reagent Installation window is updated. Go to Logs, Reagent Log, Reagent
Installation.
Autosampler - Access Door Open
The autosampler access door is open (24V interlock switch is open).
Corrective Action
1
Close the autosampler access door. The autosampler resets.
2
If the problem persists, call Siemens Service for assistance.
Summary of Contents for ADVIA 2120
Page 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Page 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Page 46: ...3 12 Daily Routine ...
Page 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Page 269: ...7 100 Status Line messages ...
Page 341: ...8 72 Methods ...