Troubleshooting the Analyzer
5-31
Replacing the fuses
IMPORTANT
All fuses must be replaced with same rated fuses.
Materials required: Flat-head screwdriver
Fuse ratings and part numbers:
FUSE APPLICATION
FUSE RATINGS
FUSE
NO.
DEVICE
PROTECTED
TYPE
100 V
120 V SYSTEMS
220 V
230 V
240 V SYSTEMS
F1
AC Input (L1)
3 AB
15.00A (T) 250V
PN625-0073-03
8.00A (T) 250V
PN625-0073-02
F2
AC Input (L2)
3 AB
15.00A (T) 250V
PN625-0073-03
8.00A (T) 250V
PN625-0073-02
F3 Computer/Monitor
at
J24 and J25
20mm
6.30A (T) 250V
PN 625-0127-11
3.15A (T) 250V
PN 625-0127-10
F4
Power Supply PS1 +5V
20mm
0.50A (T) 250V
PN625-0127-09
0.25A (T) 250V
PN 625-0127-07
F5
Power Supply PS4 +5V
Perox Lamp and
±
15V
20mm
2.00 A (T) 250V
PN 625-0127-04
1.00A (T) 250V
PN 625-0144-03
F6
Power Supply PS2
+24V
20mm
4.00A (T) 250V
PN 625-0127-05
2.00A (T) 250V
PN 625-0127-04
F7
Power Supply PS3 +5V
and
±
12V
20mm
3.15A (T) 250V
PN 625-0127-10
1.60A (T) 250V
PN 625-0127-06