8-54
Methods
Parameter Explanation
RBC Fragments
Count of RBC Fragments
RBC Ghosts
Count of RBC Ghosts
Parameter Key
Dilution Factor
The dilution factor for the PLT channel is 83.33333E-3.
PLT
Platelet count
MPV
Mean Platelet Volume
Large PLT
The count of platelets with volumes greater than 20 fL is derived from the
Platelet Volume histogram based on Integrated Analysis.
Large Platelet Count in the same units selected for PLT on the Unit Set
Configuration window of the System Setup tab.
RBC Fragments
This count includes events in the RBC Fragment area of the PLT Scatter
cytogram that have volumes less than 30fL and refractive indexes greater than
1.400.
1 RBC Ghost area
2 Platelet area
3 RBC Fragment area
The RBC Fragments Count is in the same units selected for RBC on the Unit Set
Configuration window of the System Setup tab.