Welcome to the ADVIA 2120/2120i Hematology System
1-9
•
A low-angle scatter signal corresponding to light scattered at angles between
2
°
and 3
°
A
B
C
D
E
G
I
F
H
J
A
Laser Diode Driver
Board
B
Laser Diode
C
Sample Stream
D
Hi-NA Lens
E
Beam Splitter
F
Absorption Detector
G
Asymetrical Dark Stop
H
Low Angle Scatter
Detector
I
Beam Divider
J
High Angle Scatter
Detector
Hemoglobin Colorimeter Assembly
B
A
C
A
Lamp Assembly
B
Hgb Reaction Chamber
C
UFC Assembly
The hemoglobin colorimeter assembly straddles the UFC block (
C
) at the top.
The colorimeter contains a light source (
A
) set at 3.5 Vdc and a 565 nm or 546 nm
interference filter, according to the HGB method selected.
The hemoglobin reaction chamber (
B
) is built into the UFC assembly.
Hemoglobin concentration is calculated using baseline and sample readings taken
at specific intervals during the hemoglobin sample analysis period.