7-10
Status Line messages
Possible Cause
1
Clogged
sample.
Corrective Action
Check for sample clots.
2
Insufficient
sample volume.
Sample tube
may have been
removed before
aspiration was
complete.
Re-aspirate sample.
3
Insufficient
vacuum.
Check vacuum gauge and adjust if necessary.
4
Clogged clot
filter.
Remove clot filter and discard as biohazardous material.
Replace clot filter.
5
Loose
connection in
sample line.
Check connections and tighten if necessary.
6
Clog in needle.
Perform a needle rinse. Select
Utilities
menu,
Hydraulic
Functions
tab,
Needle
Rinse
.
7
Bent or
damaged
needle.
Replace the needle.
Follow the steps in Replacing the Sampler Needles.
8
Faulty
Conductivity
Detector
Contact your Siemens Field Service representative for
assistance.
Aspiration Failed - Probe Clog - Stop
Conductivity detectors did not detect a sample within the 20 seconds of
aspiration. The number of consecutive occurrences of this condition has met the
criterion set for a stop message.
Possible Cause
1
Clogged
sample.
Corrective Action
Check for sample clots.
2
Insufficient
sample volume.
Sample tube
may have been
removed before
aspiration was
complete.
Re-aspirate sample.
3
Insufficient
vacuum.
Check vacuum gauge and adjust if necessary.
Summary of Contents for ADVIA 2120
Page 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Page 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Page 46: ...3 12 Daily Routine ...
Page 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Page 269: ...7 100 Status Line messages ...
Page 341: ...8 72 Methods ...