10-16
ADVIA Autoslide Slide Maker Stainer
BIOHAZARD
All products or objects that come in contact with human or animal body fluids should be
handled, before and after cleaning, as if capable of transmitting infectious diseases. Wear
facial protection, gloves, and protective clothing.
CAUTION
Do not move the support structure that the ADVIA Autoslide resides on more than 0.3 m
(1 ft). Make sure that the interface cable and ground strap has enough slack when you
move the support structure.
1.
Make sure all system activity is complete before you disconnect the 90°
configuration.
NOTE:
Make sure that no slides are present in the Autoslide Stainer.
2.
Place the ADVIA 2120/2120i system in Standby.
3.
Select Off on the control panel to shut off the ADVIA 2120/2120i system.
This prevents any further system function. The system displays the following
message, “communication error with autoslide”. The Autoslide icon remains green.
NOTE:
The Autoslide system can remain on.
4.
Unscrew the Austoslide sample blood line (
1
).
Figure 1. Sample Blood Line
Summary of Contents for ADVIA 2120