Methods
8-35
%EOS
The percent of eosinophils.
#EOS
The absolute count of eosinophils.
Eosinophil Count
The number of events in the
Eosinophil area (1) of the
PEROX cytogram.
%LUC
The percent of large unstained cells.
#LUC
The absolute count of large unstained cells.
Large Unstained Cell Count
The number of events in the
LUC area (1) of the
PEROX cytogram.
RawWBC
Perox Valid Cells x (PHA Cells
÷
PHA Total)
PeroxCalFactor
The sampler-specific calibration factor:
PeroxCalFactor = PEROX WBC (AS) x 0.001248
PeroxCalFactor = PEROX WBC (MCTS) x 0.001248
PeroxCalFactor = PEROX WBC (OTS) x 0.001248
To view the PEROX WBC calibration factors, select
Cal/Gain Logs
at the
System Logs menu.