Maintaining the Analyzer
4-15
CAUTION
After repositioning the aspirator assembly, finger-tighten the thumb screws,
being careful that they are not cross threaded. Overtightening of the screws
can warp the baseplate, which will cause misalignment of the sampler. Mis-
threading the thumb screws can cause needle damage.
10.
Place the tubes going to the autosampler centering collar into the "hook" on
the side of the IDee reader.
11.
Close the front cover.
12.
Run saline and whole blood primers to verify system performance.
Replacing the Sheath Filters
There are two sheath filters located on either side of the analyzer. The perox
sheath filter is in a green housing on the left side and the RBC/baso sheath filter
is in a clear housing on the right side of the instrument.
Time: 10 minutes
Materials required:
•
Perox sheath filter
PN 518-3148-05
•
RBC/baso sheath filter
PN 518-3148-06
•
paper towels
Analyzer mode: Ready to Run
Location of the perox (1) and the
RBC/baso (2) sheath filters
To replace the sheath filters
1.
Carefully pull the filter toward you just enough to
remove it from its mounting and gain access to the
tubing. Place paper towels under the filter to catch
any fluid that leaks out when you remove the filter.
2.
Disconnect the reagent line (
1
) attached to the barbed
fitting (
2
) at the input port (
3
) of the filter.
Summary of Contents for ADVIA 2120
Page 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Page 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Page 46: ...3 12 Daily Routine ...
Page 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Page 269: ...7 100 Status Line messages ...
Page 341: ...8 72 Methods ...