Regulatory information
9-63
Method Data Summary
CBC Method Data
The sample stability, expected ranges, and performance characteristics for the CBC method are listed in the following Table.
Sample
Stability
Expected
Imprecision Correlation Data
Observed
Analytical Range
Room Temp
Refrigerated
Values
Mean
SD
CV%
y=
r
Sy.x
Carryover
Range
Deviation
WBC (10³/µL)
36 h
56 h
3.8 to 8.6
5.91
0.16
2.7
1.06x - 0.74
0.997
1.42
0.4%
0.02 to 21.83
20.48 to 409.55
0.46 x 10³/µL
4.1 %
RBC (10
6
/µL)
48 h
72 h
4.1 to 6.0
4.82
0.05
1.0
0.98x + 0.11
0.997
0.07
0.2 %
0.0 to 2.65
2.65 to 6.76
0.03 x 10
6
/µL
1.1 %
HGB (g/dL)
72 h
72 h
10 to 16
13.3
0.11
0.8
1.01x - 0.2
0.997
0.2
0.2 %
0.0 to 8.6
8.6 to 22.5
0.2 g/dL
2.5 %
MCV (fL)
8 h
24 h
76 to 100
81.0
0.3
0.4
0.90x + 9.6
0.994
1.1
CH (pg)
36 h
53 h
24 to 35
26.5
0.05
0.2
1.00x + 1.2
0.990
0.6
CHCM (g/dL)
8 h
24 h
29 to 34
30.6
0.14
0.5
1.17x - 6.5
0.974
0.5
RDW (%)
72 h
72 h
12 to 15
12.3
0.07
0.6
0.86x + 1.3
0.980
0.6
HDW (g/dL)
72 h
72 h
1.9 to 3.0
2.56
0.02
0.8
0.98x + 0.06
0.959
0.13
PLT (10³/µL)
48 h
48 h
140 to 360
256
6.9
2.7
1.08x - 2.0
0.991
27
0.1 %
5 to 208
205 to 3983
10 x 10³/µL
4.9 %
MPV (fL)
8 h
8 h
7 to 9
8.1
0.11
1.4
0.76x + 0.9
0.925
0.4
Hct (%)
33 to 57
n/a
MCH (pg)
n/a
n/a
24 to 31
n/a
n/a
n/a
n/a
n/a
n/a
MCHC (g/dL)
28 to 34
n/a n/a
n/a
not available or not applicable for this parameter
x= Technicon H•3 RTC
System y= ADVIA 120 Hematology System
Summary of Contents for ADVIA 2120
Page 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Page 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Page 46: ...3 12 Daily Routine ...
Page 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Page 269: ...7 100 Status Line messages ...
Page 341: ...8 72 Methods ...