2-4
Turning the System On / Off
Turning the system off
IMPORTANT
To ensure good system performance, turn the computer off at least once a week.
You must follow the proper procedure when shutting down the ADVIA
2120/2120i Hematology System. Failure to do so can result in corruption of the
database.
Turning the system off: In an Emergency
1.
Select
Off
at the analyzer touchpad.
2.
If emergency conditions permit, select
Shut Down ADVIA
at the Log
On / Off tab.
3.
Set the main power switch to Off.
NOTE
When you turn off the analyzer in an emergency, it is unable to drain some lines
as it would normally. As a result, when you restart the system, you should
perform a System Wash, using the Hydraulic Functions tab on the Utilities menu,
before running samples.
Turning the system off: Routinely
1.
At the Routine
Operations menu, select
Log On / Off
.
2.
Select
Shut Down ADVIA
.
Wait 1–2 minutes while the software shuts down.
3.
Set the computer power switch to Off.
Exiting the ADVIA 2120/2120i software
To exit the ADVIA 2120/2120i software without turning off the computer power
1.
At the Operations menu, select
Log On / Off
.
2.
Select
Shut Down ADVIA
.
Wait 1-2 minutes while the software shuts down.
3.
To restart the software, proceed with step 2 of Turning the Computer On.
Summary of Contents for ADVIA 2120
Page 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Page 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Page 46: ...3 12 Daily Routine ...
Page 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Page 269: ...7 100 Status Line messages ...
Page 341: ...8 72 Methods ...