Status Line messages
7-57
Corrective Action
Call Siemens Service for assistance.
Invalid Numerical Result Format
A result edited in the Review and Edit window has a format incompatible with
the communications protocol.
Corrective Action
Verify that the format shows four results of one character each.
Invalid RBC or WBC Flag
The RBC or WBC has incorrect or undefined values in the analyzer. The sample
will not be transmitted to the host.
Corrective Action
Verify that the result is composed of four results of one character each.
Invalid System Number
The computer system number, as it applies to the host, is improperly defined or
undefined in the parameter file.
Corrective Action
Call Siemens Service for assistance.
Invalid Test Code in Report.par File
A test code entered in the Report.par file is not valid.
Corrective Action
Correct problem using Key list in Format window.