1-10
Welcome to the ADVIA 2120/2120i Hematology System
These voltage readings correspond to the amount of transmitted light that passes
through the reaction chamber when it contains sample mixed with reagent or
rinse.
The voltage readings are then converted to a digital form by the Hgb Interface
Board and sent to the analyzer
CPU
to calculate the optical density and derive the
Hgb concentration.
Sample, Sheath, and Diaphragm Pumps
2
1
3
4
The perox sample (
1
) and sheath (
2
) pumps are located on the left side of the analyzer.
The RBC/baso/retic sample (
3
) and sheath (
4
) pumps are located on the right side
of the analyzer.
The diaphragm pumps for sheath, rinse, and wash (five in all) are located above
the reagent containers.