Status Line messages
7-59
Left Shift - Stop
The number of consecutive occurrences of this condition has met the criterion
specified for a stop message. The autosampler halts at the same time the stop
message occurs. The specified stop criterion is user definable in the Customize
Menu, System Setup, Alarm/Stop Criteria. The counter associated with the Stop
Criteria is reset when the Start/Stop button is selected.
Length of Range Field Too Long in REPORT.PAR File
The ranges have been incorrectly defined in the Report.par file.
Corrective Action
Use the Format window to correct the problem.
M
Macrocytosis - Alarm
The number of consecutive occurrences of this condition has met the criterion
specified for an alarm message. The specified alarm criterion is user definable in
the Customize Menu, System Setup, Alarm/Stop Criteria. The counter associated
with the Alarm Criteria is automatically reset.
Macrocytosis - Stop
The number of consecutive occurrences of this condition has met the criterion
specified for a stop message. The autosampler halts at the same time the stop
message occurs. The specified stop criterion is user definable in the Customize
Menu, System Setup, Alarm/Stop Criteria. The counter associated with the Stop
Criteria is reset when the Start/Stop button is selected.
Manual Sample ID - Auto Increment Off
Auto increment feature has been turned off. (Use Operations, Manual Sample
ID.) The sample ID for manually aspirated samples will not automatically
increment by one. The sample ID must be entered by keyboard, barcode reader,
or workorder.
Manual Sample ID - Auto Increment On
Auto increment feature has been selected. (Use Operations, Manual Sample ID.)
The sample ID for manually aspirated samples will increment by one for each
succeeding aspiration.
Summary of Contents for ADVIA 2120
Page 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Page 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Page 46: ...3 12 Daily Routine ...
Page 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Page 269: ...7 100 Status Line messages ...
Page 341: ...8 72 Methods ...