9-44
Regulatory Information
CSF RBC Low Level Linearity
Pool
(% of reference level)
Observed
RBC
Expected
RBC
Absolute
Deviation
%
Deviation
0.00% 0
0
0
0.05% 2
1
1
0.1% 4
3
1
0.2% 9
6
3
1.0% 34
29
5
10.0% 308
288
20
7
100% (reference level)
2880
2880
0
0
NOTE:
Although RBC linearity has been demonstrated up to 2880 RBC/
μ
L,
samples with greater than 1500 RBC/
μ
L should be diluted to ensure the accuracy
of the white count.