ADVIA Autoslide Slide Maker Stainer
10-51
Event Text
Severity
Probable Cause
Suggested Action
Autoslide Command Error
Warning
Autoslide command error, or
EEPROM failed.
Retry the command.
Autoslide Daily Shutdown
Incomplete
Warning
An error occurred that prevented
the autoslide daily shutdown from
completing.
Check message log for errors, correct
any problems, and retry the daily
shutdown. If still not successful,
perform a Mechanical Initialization and
then retry the daily shutdown.
Autoslide Daily Startup
Incomplete
Warning
An error occurred that prevented
the autoslide daily startup from
completing.
Check message log for errors, correct
any problems, and retry the daily
startup. If still not successful, perform a
Mechanical Initialization and then retry
the daily startup
Autoslide is Processing
Slides
Warning
Attempt was made to go to a
configuration screen while slides
were being processed.
Wait until all slides are completed
before attempting to change
configuration.
Autoslide Manual Inlet
Door Cannot be Opened
Warning
The manual inlet door cannot be
opened because the stainer is not
enabled for use or the ADVIA
2120/2120i Autosampler is active.
Enable the stainer in the Autoslide
options menu or stop Autosampler if
appropriate, and try again.
Autoslide Manual Inlet
Door Failed to Open
Warning
An error occurred while attempting
to open the manual inlet door.
Check for something blocking the Stat
door, and try again.
Autoslide operation
cannot proceed as
Compressor is Not Ready
Warning
Instrument is in standby, so
compressor is off.
Take instrument out of Standby mode.
Autoslide Rejects
Command Execution
Warning
The Autoslide is unable to execute
the command at this time.
Action depends on the situation.
Generally no action is required.
Autoslide Weekly
Shutdown Incomplete
Warning
An error occurred during the
weekly shutdown.
Check message log for other errors,
correct problem, reset Autoslide, and
repeat weekly shutdown.
Autoslide Weekly Startup
Incomplete
Warning
An error occurred during the
weekly startup.
Check message log for other errors,
correct problem, reset Autoslide, and
repeat weekly startup.
Autoslide Weekly Startup
Needed
Warning
Weekly startup is now required.
Perform weekly startup.
Summary of Contents for ADVIA 2120
Page 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Page 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Page 46: ...3 12 Daily Routine ...
Page 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Page 269: ...7 100 Status Line messages ...
Page 341: ...8 72 Methods ...