Flags
6-27
Baso Noise (B-NO, NB)
Definition
The Baso Noise flag is triggered if the events counted in the Noise area of the
Baso cytogram are more than 10% of the baso signals.
The B-NO (NB) flag may cause a substitution of the WBC count.
The following rules determine if the WBCB or WBCP count is reported for a
CBC/Diff sample:
•
WBCB is the primary WBC count.
•
If B-NO or B-SAT flags are triggered, the WBCP count is reported if valid.
•
If a B-NO flag and a PX-NO or PX-SAT flag are triggered, the WBCB count
is reported with a sample/system flag.
•
WBCB is reported as the WBC count for samples with a WBC lower than
1.0 x 10³ /
μ
L.
The Noise area (1) of the Baso cytogram is determined by cluster analysis.
If histogram analysis is applied, the Noise area (2) is located between channels 0
and 8 on the y axis.
Cluster Analysis
1 noise cluster
Histogram Analysis
2 noise area
Results Flagged
WBC, WBCB, %BASO, #BASO
Corrective Action
Multiple occurrences of this flag, especially for consecutive samples, can indicate
a system problem. However, isolated instances of this flag are usually sample
related.
The following list of sample-related causes may not contain all conditions that
could cause this flag. There is no intention to associate the flag with specific
diagnoses.
Summary of Contents for ADVIA 2120
Page 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Page 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Page 46: ...3 12 Daily Routine ...
Page 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Page 269: ...7 100 Status Line messages ...
Page 341: ...8 72 Methods ...