Status Line messages
7-23
Autosampler - Flash Memory Failed, Reset Autosampler
The autosampler flash memory failed to respond to a routine command and might
be defective. An autosampler reset is required.
Corrective Action
1
Select
Off
on the analyzer touchpad.
2
Wait about 1 minute and select
On
to restart the analyzer.
3
If the problem persists, call Siemens Service for assistance.
Autosampler - Initialization Required
In diagnostic mode, the analyzer sent a position command before the appropriate
initialization was performed. For example, a position mixer command could have
been sent before a mixer initialization.
Corrective Action
Call Siemens Service for assistance.
Autosampler - Input Queue Jammed
Possible Cause
1
Rack has
jammed the
leading edge of
input-queue
blade.
Corrective Action
Remove and reset rack.
2
Queue pan
motion is
impeded.
Check for foreign objects in input-queue pan.
3
Feed motor
failed.
Call Siemens Service for assistance.
Autosampler - Input Queue Motion Denied
To prevent the collision of autosampler components, the system did not execute
the command to move the autosampler input queue.
Autosampler - Input Queue Motion Denied, Reset Autosampler
The autosampler attempted an inappropriate input queue motion. An autosampler
reset is required.
Summary of Contents for ADVIA 2120
Page 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Page 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Page 46: ...3 12 Daily Routine ...
Page 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Page 269: ...7 100 Status Line messages ...
Page 341: ...8 72 Methods ...