Status Line messages
7-37
2
Faulty barcode
reader.
See
1a, b, c
3
Window of
barcode reader
is dirty, or line
of sight is
obstructed.
a
Clean barcode reader window.
b
Make sure there is no visual obstruction between the
barcode reader window and the rack.
c
If problem persists, call Siemens Service for
assistance.
4
Rack barcode
label
improperly
placed,
obscured, or
absent.
Make sure the barcode label is correctly placed on the
rack, is clean, and meets specifications.
Autosampler - Rack Scan Failed
Barcode reader is unable to scan.
Possible Cause
1
Faulty barcode
reader cabling.
Corrective Action
a
At the analyzer touchpad, select
Off
.
b
Wait about 1 minute and select
On
to restart the
analyzer.
c
If the problem persists, call Siemens Service for
assistance.
2
Faulty barcode
reader.
See
1a, b, c
Autosampler - Shuttle Gate Open
Input queue cannot advance the rack because the shuttle gate is not closed.
Possible Cause
1
Shuttle gate
only partially
closed.
Corrective Action
a
Check for an obstruction to the shuttle gate.
b
Push the shuttle gate to check for spring resistance.
2
Faulty sensor
or sensor
wiring.
CAUTION
Call Siemens Service for assistance. Only Siemens
Service personnel are authorized to take this corrective
action:
Check sensor and associated wiring.
Summary of Contents for ADVIA 2120
Page 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Page 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Page 46: ...3 12 Daily Routine ...
Page 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Page 269: ...7 100 Status Line messages ...
Page 341: ...8 72 Methods ...