
6-50
Flags
NRBC Enumeration Histogram
Noise-Lymph Histogram
Suspect Large Plt Interference (NRLPLT, NP)
The Suspect Large Plt Interference flag is triggered if the system counts more
than 40,000 large platelets /
μ
L in the RBC/PLT channel.
Results Flagged
#NRBC, %NRBC, WBC, %NEUT, #NEUT, %LYMPH, #LYMPH, %MONO,
#MONO, %EOS, #EOS, %BASO, #BASO, %LUC, #LUC
NRBC Enumeration Histogram
Large platelets may obscure the nRBC
population as shown to the left.
Suspect Lipid Interference (NR-LPD, NL)
Definition
The Suspect Lipid Interference flag is triggered if the number of signals in the
noise region of the Basophil/Lobularity Channel exceeds a specified limit when
the system reports an NRBC count.
The NR-LPD flag is triggered when both of the following conditions occur:
•
BASO % Noise > 2
•
BASO % Noise x WBCB > 10
Summary of Contents for ADVIA 2120
Page 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Page 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Page 46: ...3 12 Daily Routine ...
Page 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Page 269: ...7 100 Status Line messages ...
Page 341: ...8 72 Methods ...