Status Line messages
7-35
bottom of an
improperly
positioned
rack.
assistance.
3
Pushpin
catches on the
bottom of the
mixer.
a
Check rack position in mixer. Adjust if necessary.
b
If the problem persists, call Siemens Service for
assistance.
4
Faulty output-
queue sensor
or wiring.
Call Siemens Service for assistance.
Autosampler - Pushpin Jammed
Pushpin is jammed and cannot extend or retract properly.
Possible Cause
1
Pushpin
catches on
bottom of
improperly
positioned
rack.
Corrective Action
a
Check and adjust rack position in mixer.
CAUTION
Call Siemens Service for assistance. Only Siemens
Service personnel are authorized to take this corrective
action.
b
Check output-queue sensor and sensor wiring.
2
Pushpin
catches on
bottom surface
of mixer
extrusion.
a
Check and adjust rack position in mixer.
CAUTION
Call Siemens Service for assistance. Only Siemens
Service personnel are authorized to take this corrective
action.
b
Check output-queue sensor and sensor wiring.
Autosampler - Rack Index Failed, Reset Autosampler
The scanner has read a valid barcode that does not correspond to the expected
rack or position numbers. This indicates that the rack has not moved to the proper
position.
Possible Cause
1
Rack
movement
problem.
Corrective Action
a
At the analyzer touchpad, select
Off
.
b
Wait about 1 minute and select
On
to restart the
analyzer.
c
If the problem persists, call Siemens Service for
Summary of Contents for ADVIA 2120
Page 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Page 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Page 46: ...3 12 Daily Routine ...
Page 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Page 269: ...7 100 Status Line messages ...
Page 341: ...8 72 Methods ...