10-40
ADVIA Autoslide Slide Maker Stainer
Performing Weekly Shutdown and Startup
The weekly shutdown provides a thorough cleaning of the fluidics circuit and
must be performed once a week. Failure to perform the shutdown procedure
regularly can result in line clogs.
TOXIC
Methanol is toxic. To avoid serious irreversible effects do not inhale, do not
contact with skin and do not swallow. Wear suitable protective clothing and
gloves. In case of accident or if you feel unwell, seek medical advice
immediately.
FLAMMABLE
Methanol is flammable. Keep away from sources of ignition - no smoking.
To perform a weekly shutdown
1.
On the ADVIA 2120/2120i software menu, select
Procedures
, then select
Clean
Autoslide
.
The Clean Autoslide window opens.
2.
Select
Weekly Shutdown
and follow the instructions displayed on the
window to perform a weekly shutdown.
3.
When the weekly shutdown completes, you must perform a weekly startup
before slide production can resume. If you want turn the system off after the
Weekly Shut Down, wait until you are ready to run more samples before
performing the Weekly Startup.
4.
At the Clean Autoslide window, select
Weekly Startup
.
5.
Follow the instructions displayed on the window to perform a weekly startup.
Cleaning the Stainer Wells and Plate
BIOHAZARD
All products or objects that come in contact with human or animal body fluids
should be handled, before and after cleaning, as if capable of transmitting
infectious diseases. Wear facial protection, gloves, and protective clothing.
The operator should follow the recommendations to prevent the transmission of
infectious agents in health-care settings as recommended for potentially
infectious specimens in
Protection of Laboratory Workers from Infectious
Disease Transmitted by Blood, Body Fluids, and Tissue
, 2d edition; Approved
Guideline (1997) Document M29-A, National Committee for Clinical Laboratory
Standards (NCCLS). This document contains complete information on user
protection and it can be used as reference material for instructions on laboratory
safety.
Summary of Contents for ADVIA 2120
Page 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Page 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Page 46: ...3 12 Daily Routine ...
Page 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Page 269: ...7 100 Status Line messages ...
Page 341: ...8 72 Methods ...