Checking the Impedance
Impedance checking is an important part of BIS monitoring because it tests the quality
of the BIS electrode connections. Make sure that the BIS sensor passes the impedance
check before starting BIS monitoring to avoid misleading BIS data. To optimize
performance, check the individual electrode status (provides a smoother trend with
decreased variability and sensitivity to artifact in addition to the overall status.
Individual electrode impedance value may display " ***", even though the overall
impedance passes).
To check the BIS impedance
Click on the BIS parameter box (if displayed);
or
1. Press the
Menu
fixed key.
2. Click on
Patient Setup
.
3. Click on
Parameters
.
4. Click on
BIS
.
5.
Click on
Check impedance
The BIS impedance menu, depicted below,
classifies the status of each electrode.
Pass
- Indicates that the impedance check
has passed based on the combined
electrode impedance values. You can exit
the impedance menu and resume normal
monitoring.
Fail
- Indicates that the impedance check
has failed. In this case, do the following:
— Press the electrodes firmly to in place to provide good contact.
BISx Alarms
Accesses the Alarm Limits table.
NOTE
:
BISx limit alarm is medium (see
page 5-3).
• BIS alarms;
ON
or
OFF
(default
OFF
)
• BIS high limit;
5
to
100
(default
100).
• BIS low limit;
0
to
95
(default 20).
• BIS archive;
OFF
,
Record, Store,
Str/Rec (
default
OFF)
Menu Item
Description
Available Settings
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...