![Dräger Infinity Delta Instructions For Use Manual Download Page 519](http://html.mh-extra.com/html/drager/infinity-delta/infinity-delta_instructions-for-use-manual_2526779519.webp)
F
I
O2
VF8
D
ELTA
/D
ELTA
XL/K
APPA
29-11
Infinity etCO
2
Microstream SmartPod
1. To clean the pod's surfaces, lightly dampen a cloth with a 70% alcohol
solution.
2. Wipe all surfaces.
Alcohol wipes may also be used. The frequency of the cleaning procedure should be
determined by hospital policy.
FiO
2
Clean the FiO
2
sensor by wiping external surfaces with a cloth lightly dampened with
a mild detergent or isopropyl alcohol. Disinfect the external surface of the FiO
2
sensor
using a cloth moistened with ethanol.
WARNING: Do not autoclave or sterilize this device.
CAUTION:
Do not spray or pour any liquid directly on the pod, accessories or
consumables.
Do not use caustic or abrasive cleaners, or harsh solvents, including
petroleum-based or acetone solutions, to clean the device.
Microstream ® etCO
2
consumables are designed for single patient use and are
not to be reprocessed. Do not attempt to clean, disinfect, sterilize or flush out
the FilterLine as this can damage the monitor
CAUTION:
Do not autoclave, gas sterilize, or irradiate the oxygen sensors. Do not
clean the sensor with chemicals other than alcohol or a mild cleaning agent.
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...