B-10
D
ELTA
/D
ELTA
XL/K
APPA
VF8
B T
ECHNICAL
D
ATA
Output high (QRS
detected):
Duration: 50 ms
Amplitude: +9 V ±5 % (nominal)
Source impedance: 5100
Ω
Rise time: <7 µs
Fall time: <1 µs
Output low (no QRS):
<0.8V @30 mA sink current
Pacemaker pulses:
Not included
User Interface
Controls:
Fixed keys and rotary knob
Alarms:
3 levels: High, Medium, Low
Screen
Type:
Thin Film Transistor Liquid Crystal
(TFT-LCD) Display, Active Matrix.
Size:
Delta:
264 mm (10.4 in) diagonal
Delta XL: 310 mm (12.2 in) diagonal
Viewing area:
Delta:
211 x 158 mm (8.3 x 6.2 in)
Delta XL: 246 x 184.5 mm (9.7 x 7.3 in)
Resolution:
Delta:
640 x 480 pixels
Delta XL: 800 x 600 pixels
Color capability:
512
Sweep speed:
6.25, 12.5, 25 and 50 mm/s ±10 %
Trend storage
Data storage:
24 hours of trended parameter information
Data resolution:
30 second sampling
Trend graphs:
1, 2, 4, 8, 12 and 24 hour display formats
Trend tables:
1, 5, 15, 30, and 60-minute display formats
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...