13-14
D
ELTA
/D
ELTA
XL/K
APPA
VF8
13 N
ON
-I
NVASIVE
B
LOOD
P
RESSURE
NBP Cuff Cannot
Deflate
Pneumatic failure.
• Check hose and cuff for obstructions
and replace, if necessary.
• If the message does not clear,
contact the hospital’s technical
personnel or DrägerService.
NBP Cuff
Deflation Error
NBP pump or valves have been
energized for longer than 2 minutes
(Adult or Pediatric mode) or 90 seconds
(Neonatal mode).
• Disconnect and reconnect cuff.
• Check hose and cuff for obstructions
and replace, if necessary.
• If the message does not clear,
contact the hospital’s technical
personnel or DrägerService.
NBP Cuff Leak
The drop in cuff pressure after the end
of the inflation cycle is too great.
• Check hose and cuff for leaks and
replace, if necessary.
• Restart the measurement. If the
message does not clear, contact the
hospital’s technical personnel.
NBP
Cannot Measure
The pulse profile is too poor to make a
reliable measurement (usually due to
persistent motion artifact).
• Check the patient and treat, if
necessary.
• Move the cuff to a limb with less
movement.
• Restart measurement. If the
message persists contact the
hospital’s technical personnel or
DrägerService.
NBP Blocked
Line
The inflation rate is too high during the
inflation cycle or the time to dump the
residual cuff pressure at the end of the
deflation cycle is too short.
• Select a different cuff.
• Check the hose and cuff for
damage.
• Restart the measurement. If the
message does not clear and contact
the hospital’s technical personnel or
DrägerService.
NBP Cannot
Zero
The monitor is unable to zero
transducer within 30 seconds from start
of NBP program, usually because of
motion artifact.
• Check the patient and treat if
necessary.
• Move cuff to a limb with less
movement.
• If the message does not clear,
contact the hospital’s technical
personnel.
NBP
Measurement
Timeout
A measurement was aborted (usually
due to motion artifact) because it lasted
longer than two minutes (adult or
pediatric) or 90 seconds (neonatal).
Repeat the measurement.
NBP No
Pulsation
Weak signal. The monitor is unable to
detect a sufficient number of pulsations
of adequate amplitude within two
minutes.
• Check the patient and treat, if
necessary.
• Check the hose and cuff.
• Check for proper size, placement of
cuff.
Message
Possible Cause
Suggested Action
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...