![Dräger Infinity Delta Instructions For Use Manual Download Page 348](http://html.mh-extra.com/html/drager/infinity-delta/infinity-delta_instructions-for-use-manual_2526779348.webp)
19-18
D
ELTA
/D
ELTA
XL/K
APPA
VF8
19
ET
CO2 (E
ND
-T
IDAL
CO2)
MONITORING
etCO
2
Adapter
Cal. Failed
The airway adapter calibration
(zeroing in room air) failed.
Make sure the adapter is properly attached to
the sensor and that its windows are clean. If the
problem persists, try another adapter.
etCO2
Adapter Cal.
Required
The sidestream measurement
mode was initiated, requiring air
way adapter cal. (zeroing in
room air) to calibrate pump.
Calibrate the new adapter.
etCO
2
Tubing
Blocked
Sidestream tubing obstructed, or
the filter is clogged.
• Clear blockage in the tubing.
• Replace the module/pod.
etCO
2
Tubing
Leak
The sidestream tubing has a
leak.
Change the tubing.
etCO
2
module
Disconnected
The etCO
2
module is not
connected to the monitor.
Check cable and connection. Replace cable if
necessary.
etCO
2
module
Incompatible
Corrupt EEPROM. Wrong
software or hardware version
• Try a new module.
• Consult the hospital’s technical personnel.
NOTE
: If the etCO
2
pod is disconnected, the etCO
2
parameter box disappears.
Message
Condition
Suggested Action
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...