26-18
D
ELTA
/D
ELTA
XL/K
APPA
VF8
26 P
ULSE
C
ONTOUR
C
ARDIAC
O
UTPUT
(P
I
CCO)
MONITORING
Quick Reference – PiCCO Setup
Available functions, present on Pulse Contour parameters setup menu, are described in
the following table.
Pulse Contour Parameters Setup Menu
Menu Item
Description
Available Settings
Show All
Parameters
Displays all the PiCCO (continuous
and discrete) parameters in a
separate window.
Not applicable
Parameters Setup
Allows user to designate which
parameter displays as
Parameter 1
,
Parameter 2
, and
Parameter 3
.
Parameter 1:
PCCI
(default)
,
PCCO, p-SV, p-SVI, p-SVR, p-
SVRI, SVV, PPV, dPmax
Parameter 2:
p-SVRI
(default)
,
PCCO, PCCI, p-SV, p-SVI, p-
SVR, p-SVRI, SVV, PPV, dPmax
Parameter 3
SVV
(default)
,
PCCO, PCCI, p-SV, p-SVI, p-
SVR, p-SVRI, PPV, dPmax
PiCCO Mini Trend
Designates which parameter will be
trended for Mini Trend.
PCCI
(default)
,
PCCO, p-SV, p-SVI, p-SVR, p-
SVRI, SVV
Catheter type
Read-only field. Displays the
catheter type in use.
NOTE
:
Other
is displayed if the
catheter that is connected requires
manual entry of an ACC value.
Selecting
Other
displays a key pad
for manual entry.
PV2013L07, PV2014L08,
PV2014L16, PV2014L22,
PV2015L20, PV2014L50, Other
Catheter Position
Selects the position of the catheter.
NOTE
: The selection
Radial artery
is only visible when the catheter
PV2014L50 is connected.
Femoral artery
,
Axillary artery
,
Brachial artery
,
Radial artery
NOTE
: When a patient is initially
admitted, this field displays ‘---’
until the catheter position is
selected.
CAUTION
: Confirm that the
selected catheter position is
accurate before starting a
measurement. Changing the
catheter position after starting a
measurement, has no effect on the
current measurements. To correct
the catheter position, discharge
and readmit the patient.
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...