C.O. M
EASUREMENT
P
ROCEDURES
VF8
D
ELTA
/D
ELTA
XL/K
APPA
15-11
To measure C.O. in manual mode
1. Press the
C.O. Start
fixed key on the hemodynamic pod or select
C.O. Start
on
the C.O. setup menu. The
READY
message appears for 30 seconds or until a
blood temperature drop is detected.
2. Immediately inject the saline solution and wait for the monitor to calculate a
C.O. value. The monitor begins to calculate a C.O. value as soon as it detects
a blood temperature drop.
If the monitor fails to detect the temperature drop caused by the injectate, the
waveform disappears after 30 seconds. An attention tone sounds, an error
message appears in the local message area, and three asterisks (
* * *)
appear in
the
Save AVG
field. Repeat steps 1 and 2 for additional measurements.
NOTE:
C.O. Start
is only available as a menu item in manual mode. As soon as you
initiate a C.O. measurement, the
C.O. Start
key is disabled and the
C.O. Start
menu item
is ghosted until a value is reported.
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...