15-10
D
ELTA
/D
ELTA
XL/K
APPA
VF8
15 C
ARDIAC
O
UTPUT
(C.O.)
To enter a computation constant
1. From the C.O. menu, click on
Comp. Constant
. A data-entry screen appears
on the menu’s right side:
2. Enter the computation constant and click on
Accept
to confirm your entry.
The manually entered computation constant is displayed to the right of the
Comp. Constant key.
C.O. Measurement Procedures
To measure C.O. in automatic mode
1. Press the
C.O. Start
fixed key on the hemodynamic pod to display the C.O.
averaging screen. A tone sounds and a
READY
message appears when the
monitor detects a stable blood temperature.
2. Inject the saline solution into the patient’s bloodstream after you see the
READY
message. A thermodilution curve appears, displaying the change in
blood temperature.
3. Repeat step 2 to take an additional measurement, making sure you wait for the
READY
message. If a temperature drop is not is detected within four minutes,
the Averaging screen closes, and you must repeat steps 1 and 2 for additional
C.O. measurements.
NOTE:
If the monitor detects that a new Edwards/Baxter, BD/Ohmeda, or Arrow
catheter is connected, the monitor automatically chooses the correct computation
constant for you. If the C.O. setup menu is displayed when that occurs, the monitor does
not update its displayed computation constant value until the next C.O. measurement is
taken. The correct value is used for the C.O. measurement, however.
NOTE:
If the
READY
message fails to appear or appears only intermittently, switch to
Manual mode and repeat step 2.
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...