Precautions
WARNING:
You must either remove the transducer before
defibrillating, or remembrane and calibrate the
transducer after defibrillating. To avoid transducer
damage, remove it from the patient during surgical
procedures.
Sensor temperatures of 43 °C (109 °F) and above for
long periods may damage the skin and could result in
severe burns. Change the monitoring site at least every
four hours and whenever the site timer expires. Check
the patient frequently if you are not using the site
timer.
CAUTION:
Verify the Auto Heater Shutdown setting with your unit manager before you
configure the site timer. If the Auto Heater Shutdown function has been enabled
in the Monitor Setups menu, the sensor heater shuts down as soon as the tpO
2
/
CO
2
site timer expires. If the Auto Heater Shutdown function has been disabled,
the heater does not shut down and could burn the patient’s skin.
Clean tpO
2
/CO
2
sensors with the pad included in the membrane kit. Do not
clean the sensor surface with alcohols.
After applying hand lotion, dry your hands thoroughly before handling the
sensor and cable because lotions containing isopropanol/propylalcohol and
alcohol can damage the sensor cable.
For accurate measurement values always use the Severinghaus correction
factor. The tpCO
2
label is used for the transcutaneous carbon dioxide
measurement values, whether this correction factor was applied or not.
Always store the sensor with the cap on to avoid drying out the membrane. If
the sensor has been in storage or if you are using a new sensor, soak the
membranes for 24 hours before use.
NOTE:
tpO
2
/CO
2
may be monitored through an MIB connection to the
Radiometer MicroGas 7650 monitor.
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...