5-8
D
ELTA
/D
ELTA
XL/K
APPA
VF8
5 A
LARMS
Upper and Lower Alarm Limits
Alarm limits should be set according to your patient’s prevailing condition within the
predefined ranges of the monitor listed in the following table.
Parameter
Predefined Alarm Range
Default
State
Default Alarm
Setting / Alarm Priority
ARR
See “Arrhythmia Setup Table” on page 9-6.
ART S/M/D
-5 to 300 mmHg (-0.7 to 40.0 kPa)
Off
S: (Adult)
Low 90 mmHg (12.0 kPa)
High 160 mmHg (21.3 kPa)
M: (Adult)
Low 60 mmHg (8.0 kPa)
High 125 mmHg (16.7 kPa)
D: (Adult)
Low 50 mmHg (6.7 kPa)
High 110 mmHg (14.7
kPa)
S: (Pediatric)
Low 50 mmHg (6.7 kPa)
High 120 mmHg (16.0 kPa)
M: (Pediatric)
Low 50 mmHg (6.7 kPa)
High 80 mmHg (10.7 kPa)
D: (Pediatric)
Low 30 mmHg (4.0 kPa)
High 80 mmHg (10.7 kPa)
S: (Neonatal)
Low 50 mmHg (6.7 kPa)
High 120 mmHg (16.0 kPa)
M: (Neonatal)
Low 40 mmHg (5.3 kPa)
High 85 mmHg (11.3 kPa)
D: (Neonatal)
Low 35 mmHg (4.7 kPa)
High 80 mmHg (10.7 kPa)
BIS (From
BISx pod only.
You cannot
control alarms
from Aspect
A2000)
10 to 100
Off
Low 10
High 100
BT/BT*
25.0 to 43.0
o
C (77.0 to 109.4
o
F)
Off
Low 34.0
o
C (93.2
o
F)
High 39.0
o
C (102.2
o
F)
CPP/CPP2/
CPP3/CPP4
-25 to 300 mmHg (-3.3 to 40.0 kPa) Off
Low 70 mmHg (9.3 kPa)
High 100 mmHg (13.3 kPa)
CVP
-5 to 300 mmHg (-0.7 to 40.0 kPa)
Off
Low 0 mmHg (0.0 kPa)
High 20 mmHg (2.7 kPa)
et DES/i DES
0 to 20 %
Off
Low 0% High 20%
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...